home / skills / microck / ordinary-claude-skills / opentrons-integration

opentrons-integration skill

/skills_all/opentrons-integration

This skill helps you design and automate Opentrons Protocol API v2 workflows for Flex and OT-2, enabling liquid handling and module control.

npx playbooks add skill microck/ordinary-claude-skills --skill opentrons-integration

Review the files below or copy the command above to add this skill to your agents.

Files (2)
SKILL.md
14.2 KB
---
name: opentrons-integration
description: "Lab automation platform for Flex/OT-2 robots. Write Protocol API v2 protocols, liquid handling, hardware modules (heater-shaker, thermocycler), labware management, for automated pipetting workflows."
---

# Opentrons Integration

## Overview

Opentrons is a Python-based lab automation platform for Flex and OT-2 robots. Write Protocol API v2 protocols for liquid handling, control hardware modules (heater-shaker, thermocycler), manage labware, for automated pipetting workflows.

## When to Use This Skill

This skill should be used when:
- Writing Opentrons Protocol API v2 protocols in Python
- Automating liquid handling workflows on Flex or OT-2 robots
- Controlling hardware modules (temperature, magnetic, heater-shaker, thermocycler)
- Setting up labware configurations and deck layouts
- Implementing complex pipetting operations (serial dilutions, plate replication, PCR setup)
- Managing tip usage and optimizing protocol efficiency
- Working with multi-channel pipettes for 96-well plate operations
- Simulating and testing protocols before robot execution

## Core Capabilities

### 1. Protocol Structure and Metadata

Every Opentrons protocol follows a standard structure:

```python
from opentrons import protocol_api

# Metadata
metadata = {
    'protocolName': 'My Protocol',
    'author': 'Name <[email protected]>',
    'description': 'Protocol description',
    'apiLevel': '2.19'  # Use latest available API version
}

# Requirements (optional)
requirements = {
    'robotType': 'Flex',  # or 'OT-2'
    'apiLevel': '2.19'
}

# Run function
def run(protocol: protocol_api.ProtocolContext):
    # Protocol commands go here
    pass
```

**Key elements:**
- Import `protocol_api` from `opentrons`
- Define `metadata` dict with protocolName, author, description, apiLevel
- Optional `requirements` dict for robot type and API version
- Implement `run()` function receiving `ProtocolContext` as parameter
- All protocol logic goes inside the `run()` function

### 2. Loading Hardware

**Loading Instruments (Pipettes):**

```python
def run(protocol: protocol_api.ProtocolContext):
    # Load pipette on specific mount
    left_pipette = protocol.load_instrument(
        'p1000_single_flex',  # Instrument name
        'left',               # Mount: 'left' or 'right'
        tip_racks=[tip_rack]  # List of tip rack labware objects
    )
```

Common pipette names:
- Flex: `p50_single_flex`, `p1000_single_flex`, `p50_multi_flex`, `p1000_multi_flex`
- OT-2: `p20_single_gen2`, `p300_single_gen2`, `p1000_single_gen2`, `p20_multi_gen2`, `p300_multi_gen2`

**Loading Labware:**

```python
# Load labware directly on deck
plate = protocol.load_labware(
    'corning_96_wellplate_360ul_flat',  # Labware API name
    'D1',                                # Deck slot (Flex: A1-D3, OT-2: 1-11)
    label='Sample Plate'                 # Optional display label
)

# Load tip rack
tip_rack = protocol.load_labware('opentrons_flex_96_tiprack_1000ul', 'C1')

# Load labware on adapter
adapter = protocol.load_adapter('opentrons_flex_96_tiprack_adapter', 'B1')
tips = adapter.load_labware('opentrons_flex_96_tiprack_200ul')
```

**Loading Modules:**

```python
# Temperature module
temp_module = protocol.load_module('temperature module gen2', 'D3')
temp_plate = temp_module.load_labware('corning_96_wellplate_360ul_flat')

# Magnetic module
mag_module = protocol.load_module('magnetic module gen2', 'C2')
mag_plate = mag_module.load_labware('nest_96_wellplate_100ul_pcr_full_skirt')

# Heater-Shaker module
hs_module = protocol.load_module('heaterShakerModuleV1', 'D1')
hs_plate = hs_module.load_labware('corning_96_wellplate_360ul_flat')

# Thermocycler module (takes up specific slots automatically)
tc_module = protocol.load_module('thermocyclerModuleV2')
tc_plate = tc_module.load_labware('nest_96_wellplate_100ul_pcr_full_skirt')
```

### 3. Liquid Handling Operations

**Basic Operations:**

```python
# Pick up tip
pipette.pick_up_tip()

# Aspirate (draw liquid in)
pipette.aspirate(
    volume=100,           # Volume in µL
    location=source['A1'] # Well or location object
)

# Dispense (expel liquid)
pipette.dispense(
    volume=100,
    location=dest['B1']
)

# Drop tip
pipette.drop_tip()

# Return tip to rack
pipette.return_tip()
```

**Complex Operations:**

```python
# Transfer (combines pick_up, aspirate, dispense, drop_tip)
pipette.transfer(
    volume=100,
    source=source_plate['A1'],
    dest=dest_plate['B1'],
    new_tip='always'  # 'always', 'once', or 'never'
)

# Distribute (one source to multiple destinations)
pipette.distribute(
    volume=50,
    source=reservoir['A1'],
    dest=[plate['A1'], plate['A2'], plate['A3']],
    new_tip='once'
)

# Consolidate (multiple sources to one destination)
pipette.consolidate(
    volume=50,
    source=[plate['A1'], plate['A2'], plate['A3']],
    dest=reservoir['A1'],
    new_tip='once'
)
```

**Advanced Techniques:**

```python
# Mix (aspirate and dispense in same location)
pipette.mix(
    repetitions=3,
    volume=50,
    location=plate['A1']
)

# Air gap (prevent dripping)
pipette.aspirate(100, source['A1'])
pipette.air_gap(20)  # 20µL air gap
pipette.dispense(120, dest['A1'])

# Blow out (expel remaining liquid)
pipette.blow_out(location=dest['A1'].top())

# Touch tip (remove droplets on tip exterior)
pipette.touch_tip(location=plate['A1'])
```

**Flow Rate Control:**

```python
# Set flow rates (µL/s)
pipette.flow_rate.aspirate = 150
pipette.flow_rate.dispense = 300
pipette.flow_rate.blow_out = 400
```

### 4. Accessing Wells and Locations

**Well Access Methods:**

```python
# By name
well_a1 = plate['A1']

# By index
first_well = plate.wells()[0]

# All wells
all_wells = plate.wells()  # Returns list

# By rows
rows = plate.rows()  # Returns list of lists
row_a = plate.rows()[0]  # All wells in row A

# By columns
columns = plate.columns()  # Returns list of lists
column_1 = plate.columns()[0]  # All wells in column 1

# Wells by name (dictionary)
wells_dict = plate.wells_by_name()  # {'A1': Well, 'A2': Well, ...}
```

**Location Methods:**

```python
# Top of well (default: 1mm below top)
pipette.aspirate(100, well.top())
pipette.aspirate(100, well.top(z=5))  # 5mm above top

# Bottom of well (default: 1mm above bottom)
pipette.aspirate(100, well.bottom())
pipette.aspirate(100, well.bottom(z=2))  # 2mm above bottom

# Center of well
pipette.aspirate(100, well.center())
```

### 5. Hardware Module Control

**Temperature Module:**

```python
# Set temperature
temp_module.set_temperature(celsius=4)

# Wait for temperature
temp_module.await_temperature(celsius=4)

# Deactivate
temp_module.deactivate()

# Check status
current_temp = temp_module.temperature  # Current temperature
target_temp = temp_module.target  # Target temperature
```

**Magnetic Module:**

```python
# Engage (raise magnets)
mag_module.engage(height_from_base=10)  # mm from labware base

# Disengage (lower magnets)
mag_module.disengage()

# Check status
is_engaged = mag_module.status  # 'engaged' or 'disengaged'
```

**Heater-Shaker Module:**

```python
# Set temperature
hs_module.set_target_temperature(celsius=37)

# Wait for temperature
hs_module.wait_for_temperature()

# Set shake speed
hs_module.set_and_wait_for_shake_speed(rpm=500)

# Close labware latch
hs_module.close_labware_latch()

# Open labware latch
hs_module.open_labware_latch()

# Deactivate heater
hs_module.deactivate_heater()

# Deactivate shaker
hs_module.deactivate_shaker()
```

**Thermocycler Module:**

```python
# Open lid
tc_module.open_lid()

# Close lid
tc_module.close_lid()

# Set lid temperature
tc_module.set_lid_temperature(celsius=105)

# Set block temperature
tc_module.set_block_temperature(
    temperature=95,
    hold_time_seconds=30,
    hold_time_minutes=0.5,
    block_max_volume=50  # µL per well
)

# Execute profile (PCR cycling)
profile = [
    {'temperature': 95, 'hold_time_seconds': 30},
    {'temperature': 57, 'hold_time_seconds': 30},
    {'temperature': 72, 'hold_time_seconds': 60}
]
tc_module.execute_profile(
    steps=profile,
    repetitions=30,
    block_max_volume=50
)

# Deactivate
tc_module.deactivate_lid()
tc_module.deactivate_block()
```

**Absorbance Plate Reader:**

```python
# Initialize and read
result = plate_reader.read(wavelengths=[450, 650])

# Access readings
absorbance_data = result  # Dict with wavelength keys
```

### 6. Liquid Tracking and Labeling

**Define Liquids:**

```python
# Define liquid types
water = protocol.define_liquid(
    name='Water',
    description='Ultrapure water',
    display_color='#0000FF'  # Hex color code
)

sample = protocol.define_liquid(
    name='Sample',
    description='Cell lysate sample',
    display_color='#FF0000'
)
```

**Load Liquids into Wells:**

```python
# Load liquid into specific wells
reservoir['A1'].load_liquid(liquid=water, volume=50000)  # µL
plate['A1'].load_liquid(liquid=sample, volume=100)

# Mark wells as empty
plate['B1'].load_empty()
```

### 7. Protocol Control and Utilities

**Execution Control:**

```python
# Pause protocol
protocol.pause(msg='Replace tip box and resume')

# Delay
protocol.delay(seconds=60)
protocol.delay(minutes=5)

# Comment (appears in logs)
protocol.comment('Starting serial dilution')

# Home robot
protocol.home()
```

**Conditional Logic:**

```python
# Check if simulating
if protocol.is_simulating():
    protocol.comment('Running in simulation mode')
else:
    protocol.comment('Running on actual robot')
```

**Rail Lights (Flex only):**

```python
# Turn lights on
protocol.set_rail_lights(on=True)

# Turn lights off
protocol.set_rail_lights(on=False)
```

### 8. Multi-Channel and 8-Channel Pipetting

When using multi-channel pipettes:

```python
# Load 8-channel pipette
multi_pipette = protocol.load_instrument(
    'p300_multi_gen2',
    'left',
    tip_racks=[tips]
)

# Access entire column with single well reference
multi_pipette.transfer(
    volume=100,
    source=source_plate['A1'],  # Accesses entire column 1
    dest=dest_plate['A1']       # Dispenses to entire column 1
)

# Use rows() for row-wise operations
for row in plate.rows():
    multi_pipette.transfer(100, reservoir['A1'], row[0])
```

### 9. Common Protocol Patterns

**Serial Dilution:**

```python
def run(protocol: protocol_api.ProtocolContext):
    # Load labware
    tips = protocol.load_labware('opentrons_flex_96_tiprack_200ul', 'D1')
    reservoir = protocol.load_labware('nest_12_reservoir_15ml', 'D2')
    plate = protocol.load_labware('corning_96_wellplate_360ul_flat', 'D3')

    # Load pipette
    p300 = protocol.load_instrument('p300_single_flex', 'left', tip_racks=[tips])

    # Add diluent to all wells except first
    p300.transfer(100, reservoir['A1'], plate.rows()[0][1:])

    # Serial dilution across row
    p300.transfer(
        100,
        plate.rows()[0][:11],  # Source: wells 0-10
        plate.rows()[0][1:],   # Dest: wells 1-11
        mix_after=(3, 50),     # Mix 3x with 50µL after dispense
        new_tip='always'
    )
```

**Plate Replication:**

```python
def run(protocol: protocol_api.ProtocolContext):
    # Load labware
    tips = protocol.load_labware('opentrons_flex_96_tiprack_1000ul', 'C1')
    source = protocol.load_labware('corning_96_wellplate_360ul_flat', 'D1')
    dest = protocol.load_labware('corning_96_wellplate_360ul_flat', 'D2')

    # Load pipette
    p1000 = protocol.load_instrument('p1000_single_flex', 'left', tip_racks=[tips])

    # Transfer from all wells in source to dest
    p1000.transfer(
        100,
        source.wells(),
        dest.wells(),
        new_tip='always'
    )
```

**PCR Setup:**

```python
def run(protocol: protocol_api.ProtocolContext):
    # Load thermocycler
    tc_mod = protocol.load_module('thermocyclerModuleV2')
    tc_plate = tc_mod.load_labware('nest_96_wellplate_100ul_pcr_full_skirt')

    # Load tips and reagents
    tips = protocol.load_labware('opentrons_flex_96_tiprack_200ul', 'C1')
    reagents = protocol.load_labware('opentrons_24_tuberack_nest_1.5ml_snapcap', 'D1')

    # Load pipette
    p300 = protocol.load_instrument('p300_single_flex', 'left', tip_racks=[tips])

    # Open thermocycler lid
    tc_mod.open_lid()

    # Distribute master mix
    p300.distribute(
        20,
        reagents['A1'],
        tc_plate.wells(),
        new_tip='once'
    )

    # Add samples (example for first 8 wells)
    for i, well in enumerate(tc_plate.wells()[:8]):
        p300.transfer(5, reagents.wells()[i+1], well, new_tip='always')

    # Run PCR
    tc_mod.close_lid()
    tc_mod.set_lid_temperature(105)

    # PCR profile
    tc_mod.set_block_temperature(95, hold_time_seconds=180)

    profile = [
        {'temperature': 95, 'hold_time_seconds': 15},
        {'temperature': 60, 'hold_time_seconds': 30},
        {'temperature': 72, 'hold_time_seconds': 30}
    ]
    tc_mod.execute_profile(steps=profile, repetitions=35, block_max_volume=25)

    tc_mod.set_block_temperature(72, hold_time_minutes=5)
    tc_mod.set_block_temperature(4)

    tc_mod.deactivate_lid()
    tc_mod.open_lid()
```

## Best Practices

1. **Always specify API level**: Use the latest stable API version in metadata
2. **Use meaningful labels**: Label labware for easier identification in logs
3. **Check tip availability**: Ensure sufficient tips for protocol completion
4. **Add comments**: Use `protocol.comment()` for debugging and logging
5. **Simulate first**: Always test protocols in simulation before running on robot
6. **Handle errors gracefully**: Add pauses for manual intervention when needed
7. **Consider timing**: Use delays when protocols require incubation periods
8. **Track liquids**: Use liquid tracking for better setup validation
9. **Optimize tip usage**: Use `new_tip='once'` when appropriate to save tips
10. **Control flow rates**: Adjust flow rates for viscous or volatile liquids

## Troubleshooting

**Common Issues:**

- **Out of tips**: Verify tip rack capacity matches protocol requirements
- **Labware collisions**: Check deck layout for spatial conflicts
- **Volume errors**: Ensure volumes don't exceed well or pipette capacities
- **Module not responding**: Verify module is properly connected and firmware is updated
- **Inaccurate volumes**: Calibrate pipettes and check for air bubbles
- **Protocol fails in simulation**: Check API version compatibility and labware definitions

## Resources

For detailed API documentation, see `references/api_reference.md` in this skill directory.

For example protocol templates, see `scripts/` directory.

Overview

This skill integrates Opentrons Protocol API v2 capabilities for Flex and OT-2 robots into automated pipetting workflows. It helps you write Python protocols, manage labware and deck layouts, and control hardware modules such as heater-shaker, thermocycler, and magnetic/temperature modules. The goal is reliable, testable protocols for liquid handling, multi-channel operations, and complex patterns like serial dilutions and PCR setup.

How this skill works

The skill provides patterns and code examples to define protocol metadata, load instruments and labware, and run the run(protocol: ProtocolContext) entrypoint. It exposes common liquid handling primitives (aspirate, dispense, transfer, distribute, consolidate), multi-channel handling, well/location accessors, and flow-rate tuning. Module control APIs are included for temperature, magnetic, heater-shaker, and thermocycler operations, plus utilities for simulation, delays, comments, and liquid tracking.

When to use it

  • Authoring Protocol API v2 scripts for Flex or OT-2 robots
  • Automating liquid transfers, serial dilutions, plate replication, or PCR setup
  • Controlling hardware modules (thermocycler, heater-shaker, magnetic, temperature)
  • Optimizing tip usage and multi-channel pipetting for 96-well workflows
  • Simulating and validating protocols before running on the robot

Best practices

  • Set metadata apiLevel to the latest stable version
  • Label labware and define liquids for clearer logs and tracking
  • Simulate protocols first and validate deck layout to avoid collisions
  • Plan tip usage and use new_tip='once' where safe to conserve tips
  • Adjust flow rates and use air gaps/mix steps for viscous or volatile reagents
  • Add protocol.comment() and pause() points for debugging and manual interventions

Example use cases

  • Write a serial dilution across a plate with automated mixing and tip control
  • Replicate whole 96-well plates between source and destination with multi-channel pipettes
  • Prepare PCR reactions on a thermocycler module: distribute master mix, add samples, and run profile
  • Perform magnetic bead separations using the magnetic module and timed incubations
  • Run temperature-controlled incubations on heater-shaker with shaking and temperature profiles

FAQ

Which API level should I target?

Use the latest stable Protocol API v2 supported by your robot and specify it in metadata to ensure compatibility.

How do I avoid running out of tips mid-protocol?

Calculate required tips up front, include enough tip racks on the deck, and use tip reuse strategies (new_tip='once') only when appropriate for contamination risk.